Service Manual Functional Verification 33
Select Main Menu > System Settings > Device Info > Service Info > Serial
Number.
2. If the main board was replaced and a configuration files were exported, import the
configuration files, see “Import Files” on page 24.
3. Print the Configuration of the unit.
Select Main Menu > System Settings > Device Info > Print Settings
4. Verify the settings from the printout.
a. Check for software version installed.
b. S/N of the unit.
c. Interpretation (if applicable).
If imported, check any customized customer settings.
SD Slot Test
1. Insert a formatted SD memory card into the unit.
2. Choose Main Menu > System Settings > Device info > Service Info > Service
Tests>MMCSDTEST.EXE file.
3. Press G to start the test. After 30 seconds the MyFile.TXT increases one increment.
4. Press S to stop the test and confirm that no errors are reported.
ECG Communication Test (ECG Input test and lead off)
Test the patient cable and lead off indications connection.
1. Connect a known good Patient Cable w/leads
2. Connect to ECG simulator with V1/C1 lead disconnected.
3. Power ON unit.
4. Verify the lead placement screen shows V1/C1 lead flashing and all other leads
should appear not flashing.
5. Remove LL from the Simulator and verify LL flashes with all V/C leads grayed, RA,LA
& RL should not be flashing.
6. Connect all leads to the simulator.
7. Verify all leads stop flashing and the next screen appears in approximately three
seconds.
Spirometer Communication Test
CP 200 (Only)
Verify and test the connection of a Spirometer Dongle to the serial port.
1. Attach a known good spirometer dongle to the spirometer port on the side of the unit.
2. Power ON unit.