34 Functional Verification Welch Allyn CP 100 and CP 200 Electrocardiograph
3. Press the Spirometry button
4. Verify that the unit opens the Spirometry module without errors.
USB Communications Test
Verify the test connection between the unit and the CardioPerfect Workstation computer.
1. Connect the unit to the computer using a USB A-A Cable.
2. The computer should recognize the CP 200 connected and load the driver for
communication.
3. Transfer an ECG record to the CardioPerfect Workstation.
a. Press the Menu button.
b. Choose Test Directory.
c. Press the Search softkey.
d. Press the WorkStation softkey.
4. Verify the record was transferred into the workstation.
Beeper Test
Turn on and test the Audible beeper by using the keyboard keys.
1. Po we r On unit.
2. Choose Main Menu > System Settings > Device Configuration >
Audio Beeper > set to ON.
3. Exit the Main Menu
4. Verify that an Audible beep is heard when invalid keys are pressed on the keyboard.
5. If the Audio Beeper was originally set to off, choose Main Menu >
System Settings > Device Configuration > Audio Beeper > set to OFF to return
the Audio Beeper to its original setting.

Table of Contents

Other manuals for Welch Allyn CP 100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn CP 100 and is the answer not in the manual?

Welch Allyn CP 100 Specifications

General IconGeneral
BrandWelch Allyn
ModelCP 100
CategoryMedical Equipment
LanguageEnglish

Related product manuals