Service Manual CP 100 and CP 200 Electrocardiograph 63
Replacing the IEC (AC input) Connector
Procedure
1. Open unit using the Disassembly
Procedures (page 49) directions.
2. Remove printer assembly using the
Replacing the Printer Assembly (page 50)
directions.
3. Disconnect connections from the back of
the IEC connector.
4. Press in locking devices located on each
side of the IEC connector.
5. While pressing in locking devices, apply
pressure out on the connector.
Note Do one side at a time.
6. After the locking devices are released, pull
the connector out of the housing. See
Figure 53.
Figure 53. IEC (AC input) Connector
IEC
Connector
Housing