60 Disassembly and Repair Procedures Welch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Battery (DC) Fuse
If the battery (DC) fuse requires replacing, as described in “Replacing the Battery or Fuse
on page 58. For fuse value, see “Fuses” on page 77.
Procedure
1. Remove and discard the fuse. See
Figure 50.
You may need to pull hard. Use pliers
gently if necessary.
2. Connect a new fuse. It goes in either way.
3. Go to Step 7 on page 59.
Figure 50. Battery Fuse Remove