C
83
Software Updates
.
Caution Do not unplug or turn unit off during this process. It will cause the
system to be inoperable.
Note
You cannot uninstall updates made to the unit.
Procedure Figure 59. CP 200 “Software Upgrade” Screen
1. Power ON unit.
2. Insert the SD memory card into the slot on the unit.
3. Press Menu key.
4. Choose Main Menu > Systems Settings> Device Info >
Upgrade Software
The Upgrade Software screen appears. See Figure 59.
5. Press the “Upgrade” Softkey button.
The upgrade takes approximately 60 seconds to
complete.The unit will reboot.
6. Reset the Date and Time on the unit after it reboots. See the
Directions for Use manual.
Verify Update
1. Choose Main Menu > System Settings > Device Info >
Print Setting.
2. Compare product version to what was installed.
Upgrade Software
Currently Installed
On Memory Card
X.X.X
X.X.X
Upgrade