36 Functional Verification Welch Allyn CP 100 and CP 200 Electrocardiograph
Table 2. Print Verification Check List
Figure 29. Test Printout Example
Point Description
Point A Verify the distance of the arc on the circle from the top edge of the paper is 37mm +/-
3mm
Point B Verify the distance of the arc on the circle from the left edge of the paper is 45mm +/-
5mm
Point C Verify line is straight and without breaks from top to bottom
Point D Verify thick black line is Uniform in Blackness without breaks from top to bottom
Point E Verify line is straight and measures 200mm +/- 4mm
Point F Verify the print is readable
Point G Verify the lead in and output trail of the cue sensor are between 19 and 82mm down
from the 200mm line measured in point E
Point H Verify the cue sense pattern is between 97 and 135mm down from the 200mm line
measured in point E