EasyManua.ls Logo

Welch Allyn VSM 6000 series - Page 123

Welch Allyn VSM 6000 series
143 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Secondary label Select a secondary identification label for patients.
A secondary label displays only on the Home tab,
after the primary label.
Require patient ID to save readings Select this option to make entering a patient ID a
prerequisite for saving measurements. If they fail to
enter an identifier, the monitor prompts them when
they try to save.
Search by patient ID Select this option to enable clinicians to enter a
patient ID to query for the patient's information. If
clinicians scan the ID onto the Home tab, the
monitor queries the patient list and the network. If
clinicians enter the ID on the Summary tab, the
monitor queries the patient list. In both cases,
returned patient information populates the Patient
frame on the Home tab and fields on the Summary
tab.
5. Do one of the following:
To continue in the Advanced Settings, touch another tab.
To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify clinician ID settings
Clinician identification appears next to the medicine symbol in the Device Status area on the Home
tab.
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Data Management tab.
3. Touch the Clinician IDs tab.
4. Specify settings.
Setting Action/Description
Label Select a type of clinician identification label for
display on the Home tab: Full name,
Abbreviation, Clinician ID, or Symbol only
Require clinician ID to save readings Select this option to make entering a clinician ID a
prerequisite for saving measurements. If they fail to
enter identification, the monitor prompts them
when they try to save. Clinicians can enter clinician
identification on the Clinician tab.
5. Do one of the following:
To continue in the Advanced Settings, touch another tab.
To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify clinical data settings
1. Access the Advanced Settings.
Directions for use Advanced settings 117