Maintenance and service
Perform periodic checks
Welch Allyn recommends that each facility conduct periodic checks of each monitor.
1. Check the following at least daily:
• Audio (speaker and piezo beeper tones), especially at startup
• Fan, especially at startup
• Touchscreen alignment
• Date
• Time
2. Visually inspect the following at least weekly:
• the monitor for any damage or contamination
• all cables, cords, and connector ends for damage or contamination
• all mechanical parts, including covers, for integrity
• all safety-related labeling for legibility and adhesion to the monitor
• all accessories (cuffs, tubing, probes, sensors) for wear or damage
• documentation for current revision of the monitor
3. Visually inspect the following at least monthly:
• Mobile stand wheels for wear and faulty operation
• Mounting screws on wall or cart for looseness and wear
Update settings, replace items, or call for service as necessary based on results of visual
inspection. Do not use the monitor if you see any signs of damage. Qualified service personnel
must check any monitor that is damaged for proper operation before putting the monitor back into
use.
Replace the printer paper
The printer is located on the top of the monitor. Follow these steps to replace the roll of printer
paper:
1. Grasp the two tabs and pull up to open the printer door.
2. Remove the empty core.
87