3. Touch the Demo tab.
4. Specify settings.
Setting Action/Description
Type Select a type of demonstration mode.
Start Touch Start to put the monitor in demonstration
mode. Navigate to the Home tab to begin Demo
mode.
5. Do one of the following:
• To continue in the Advanced Settings, touch another tab.
• To exit the Demo mode, touch Exit on the Home tab. The monitor restarts automatically.
Parameters
Specify advanced NIBP settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Parameters tab.
3. Touch the NIBP tab.
4. Specify settings.
Setting Action/Description
Default view Select primary and secondary views.
Select Display MAP to display mean arterial
pressure (MAP) in the NIBP frame on the Home tab.
If Display MAP is selected, specify which
numerics are primary in the NIBP frame. On the
Home tab, clinicians can touch the NIBP frame to
toggle between views.
Default patient type Select a default patient type for this monitor. The
patient type shows in the Patient frame on the
Home tab.
In the Patients tab on the Summary tab, clinicians
can change the displayed patient type from the
default patient type that you set here.
Tube type Select the number of tubes that are connected to
the NIBP cuff that is used with this monitor. If you
select 1 tube, the only algorithm available for
selection is Step.
Unit of measure Select the NIBP unit of measure for display.
Algorithm and Cuff inflation target (CIT) Select the default algorithm used to determine
NIBP measurements.
Directions for use Advanced settings 113