Specifications
Physical specifications
Protection classifications, all monitor configurations
Characteristic Specification
Electrical rating 100 – 240 V AC, 50 – 60 Hz, 1.5 – 0.8 A
Duty cycle Continuous operation
Type of protection against electric
shock
Class I equipment (protectively earthed) with double insulation
Degree of protection against electric
shock, for parts applied to patients
Type BF defibrillator proof
IEC EN 60601-1, 2nd Edition
Recovery time following defibrillator
discharge
Less than or equal to 10 seconds
Flammable anesthetics
WARNING Not suitable for use with flammable
anesthetics.
Degree of protection provided by the
enclosure with respect to harmful
ingress of liquids
IPX0
Non-protected according to EN/IEC 60529; Pulse oximeter equipment
complies with ISO 9919 Cl. 44.6 Ingress of liquids tests and EN/IEC
60601-1, 60601-2-30, 60601-2-49 Cl. 44.3 Spillage tests
Height 10 in. (25.4 cm )
Width 11 in. (29.2 cm)
Depth 6 in. (15.7 cm)
Weight (including battery) 9.5 lb. (4.3 kg)
Graphical display resolution
Display area 8 in. (H) x 4 in. (V) (19.5 [H] cm x 11.3 [V] cm)
Pixels 1024 (H) x 600 (V)