What to do if... Multi 3430
102
ba75865e08 06/2010
No stable measured
value
Obviously incorrect
measured values
6.3 Dissolved oxygen
Error message,
OFL
Cause Remedy
IDS pH sensor:
Junction contaminated Clean junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if neces-
sary
IDS pH sensor + test sample:
Conductivity too low Use suitable IDS pH sensor
Temperature too high Use suitable IDS pH sensor
Organic liquids Use suitable IDS pH sensor
Cause Remedy
IDS pH sensor:
IDS pH sensor unsuitable Use suitable IDS sensor
Temperature difference
between buffer and test sample
too great
Adjust temperature of buffer
or sample solutions
Measurement procedure not
suitable
Follow special procedure
Cause Remedy
Measured value outside the
measuring range
Select a suitable measuring
medium