Operation Multi 3430
54
ba75865e08 06/2010
Continuing with two-
point calibration
4.6.8 Displaying calibration records
The calibration data can be displayed and then output to the interface.
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the <CAL
_> key.
The calibration records of the last 10 calibrations are to be found in the
menu, Calibration / Calibration data storage. To open the Calibration
menu, press the <MENU/ENTER> key in the measured value display.
9 Thoroughly rinse the IDS pH sensor with distilled water.
10 Immerse the IDS pH sensor in buffer solution 2.
11 Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The measured parameter flashes.
12 Wait for the measurement with stability control to be com-
pleted.
Set the nominal buffer value for the measured temperature with
<S><T>.
13 Take over the calibration value for the slope with <MENU/
ENTER>.
The calibration is completed as a two-point calibration.
The calibration record is displayed.