Operation Multi 3430
64
ba75865e08 06/2010
4.8.2 Measuring
Selecting the
displayed
measured parameter
You can switch between the following displays with <M>:
z D.O. concentration [mg/l]
z D.O. saturation [%]
z D.O. partial pressure [mbar].
Salinity correction When measuring the concentration of solutions with a salt content of
more than 1 g/l, a salinity correction is required.
You have the following options to determine the salinity of solution:
z Salinity measurement via an IDS conductivity sensor.
When the measured salinity value is taken from an IDS conductivity
sensor, the [Sal
] status indicator is displayed in the measuring win-
dow of the D.O. sensor. In the measuring window of the IDS conduc-
tivity sensor providing the measured salinity value the [Sal
] status
indicator is displayed.
z Manual determination and input of the salinity.
Note
You can switch the salinity correction on or off and enter the salinity in
the menu for calibration and measurement settings (see section 4.9.4).
Freezing the measured
value (HOLD function)
With the HOLD function, you can freeze the current measured value.
The displayed measured value stops changing until you switch the
HOLD function off.
1 Perform the preparatory activities according to section 4.9.1.
2 Immerse the D.O. sensor in the test sample.
30.09.2009 08:00Info
1 Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.

Table of Contents

Other manuals for wtw Multi 3430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 3430 and is the answer not in the manual?

wtw Multi 3430 Specifications

General IconGeneral
Brandwtw
ModelMulti 3430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals