Operation Multi 3430
94
ba75865e08 06/2010
4.12 Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.12.1 Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH The following settings for pH measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the menu
for calibration and measurement settings. To open the settings, display
the required parameter in the measured value display and press the
<MENU/ENTER> key.
ORP The following settings for ORP measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the menu
for calibration and measurement settings. To open the settings, display
Setting Default settings
Buffer AutoCalTEC
Cal. interval 7 d
Unit for slope mV/pH
Measured parameter pH
pH resolution 0.001
mV resolution 0.1
Asymmetry 0 mV
Slope -59.2 mV
Man. temperature 25 °C
Single-point calibration Off
Setting Default settings
mV resolution 0.1
Man. temperature 25 °C