Multi 3430 Operation
53
ba75865e08 06/2010
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
6 Wait for the measurement with stability control to be com-
pleted.
Set the nominal buffer value for the measured temperature with
<S><T>.
7 Take over the calibration value for the asymmetry with <MENU/
ENTER>.
The calibration display for the next buffer appears.
8 If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.
30.09.2009 08:00
30.09.2009 08:00