Operation Multi 3430
40
ba75865e08 06/2010
Calibration /Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the
Nernst slope of
-59.2 mV/pH (100 x deter-
mined slope/Nernst slope).
QSC /First calibra-
tion
- Starts the initial calibration
with QSC buffers.
This menu item is only avail-
able as long as no initial cal-
ibration was carried out with
the connected IDS sensor.
QSC /Record of first
calibration
- Displays the calibration
record of the QSC initial cal-
ibration
QSC /Control cali-
bration
- Starts the control calibration
with QSC buffers.
This menu item is only avail-
able if an initial calibration
was carried out with the con-
nected IDS sensor.
Man. temperature -25 ... +25 ...
+130 °C
Entry of the manually deter-
mined temperature
This menu item is only
available if an IDS adapter
and is connected.
Alternative tempera-
ture
on
off
Takes the measured tempe-
rature value from an IDS
sensor.
This menu item is only
available if an IDS adapter
and an IDS sensor with tem-
perature sensor are con-
nected.
Menu item Possible set-
ting
Description