EasyManuals Logo

wtw Multi 3430 User Manual

wtw Multi 3430
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
Multi 3430 Operation
61
ba75865e08 06/2010
Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values. The display of the measured
parameter flashes until a stable measured value is available.
You can start the Stability control manually at any time, irrespective of
the setting for automatic Stability control (see section 4.3.3) in the Sys-
tem menu.
Note
You can prematurely terminate the Stability control function manually
with <MENU/ENTER> at any time. If the Stability control function is pre-
maturely terminated, the current measurement data are output to the
interface without the AutoRead info.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
3 Select the mV display with <M>.
1 Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
2 Using <MENU/ENTER>, activate the Stability control function
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized. The progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Mea-
surement data meeting the stability control criterion is marked
by AR.
3 Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back
to the previous indication.
Measured param-
eter
Time interval Stability in the time inter-
val
ORP voltage 15 seconds
Δ mV: Better than 0.3

Table of Contents

Other manuals for wtw Multi 3430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 3430 and is the answer not in the manual?

wtw Multi 3430 Specifications

General IconGeneral
Brandwtw
ModelMulti 3430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals