Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
Measured
parameter
Time interval Stability in the time interval
D.O. concen-
tration
20 seconds
Δ : better than 0.03 mg/l
D.O. saturation 20 seconds
Δ : better than 0.4 %
D.O. partial
pressure
20 seconds
Δ : Better than 0.8 mbar