Operation Multi 3430
80
ba75865e08 06/2010
Type Measuring cell used
Cal Measuring cells whose cell con-
stant is determined by calibration
in the KCL control standard solu-
tion.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant
is displayed in the status line.
man Freely adjustable cell constant in
the range 0.450 ... 0.500 cm
-1
.
Man. cell const. 0.450 ...
0.475 ...
0.500 cm
-1
The cell constant and be dis-
played and set here
Temp. comp. (TC)
/ Method
nLF
lin
off
Procedure for temperature com-
pensation (see section 4.9.3).
This setting is only available for
the measured parameters, ϰ and
ρ.
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Coefficient of the linear tempera-
ture compensation.
This setting is only available
when the linear temperature
compensation is set.
Temp. comp. (TC)
/ Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for
the measured parameters, ϰ and
ρ.
TDS factor 0.40 ... 1.00 Factor for TDS value
Reset - Resets all sensor settings to the
delivery condition (see section
4.12.1)
Menu item Possible
setting
Description

Table of Contents

Other manuals for wtw Multi 3430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 3430 and is the answer not in the manual?

wtw Multi 3430 Specifications

General IconGeneral
Brandwtw
ModelMulti 3430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals