EasyManua.ls Logo

wtw Multi 3430 - Page 83

wtw Multi 3430
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multi 3430 Operation
83
ba75865e08 06/2010
Determining the cell
constant
For this calibration procedure, the Type setting must be set to cal . Pro-
ceed as follows to determine the cell constant:
Display Calibration
record
Cell constant [cm
-1
]
+++ Within the range
0.450 ... 0.500 cm
-1
Error Error
Outside the range
0.450 ... 0.500 cm
-1
Eliminate the error according to
chapter 6 W
HAT TO DO IF...
1 Connect an IDS conductivity sensor to the meter.
2 In the measured value display, select the conductivity parame-
ter with <M>.
3 Start the calibration with <CAL>.
The cell constant that was calibrated last is displayed.
4 Immerse the IDS conductivity sensor in the control standard
solution, 0.01 mol/l KCI.
5 Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
30.09.2009 08:00

Table of Contents

Other manuals for wtw Multi 3430

Related product manuals