For details on the setup screen that appears when you select a setting and the descriptions of the settings, 
see section 2.3.
File (FILE) Conditions
4.
  Tap the File tab to display the following screen.
   
Set the report format.
See page 2-25.
Set the file name.
See page 2-24.
Set the file report 
format.
See page 2-26.
File tab
Set comments for file names. 
See page 2-27.
Set the label 
content.
See page 2-27.
method.
See page 2-24.
Initializes the instrument’s internal memory (inside the USER/DATA folder).
Note
For details on the setup screen that appears when you select a setting and the descriptions of the settings, 
see section 2.4.
Executing an Averaged Measurement
5.
 Tap Wavelength to set the wavelength.
   
6.
 Press AVG. A measurement will start.
Measurements are performed four times using the set averaging conditions. A wait screen is displayed 
while the measurement is in progress.
Also, during measurement, a mark appears at the top of the display to indicate that the laser light is on. 
When averaged measurements are completed, measurement is stopped automatically, event analysis is 
executed, and the analysis results are displayed using icons on the screen. If you press AVG again during 
averaged measurement, the measurement is aborted. When you abort the measurement, the measurement 
results are not displayed.