EasyManuals Logo
Home>YOKOGAWA>Recording Equipment>GX20

YOKOGAWA GX20 User Manual

YOKOGAWA GX20
707 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #663 background imageLoading...
Page #663 background image
5
All Rights Reserved. Copyright © 2014, Yokogawa Electric Corporation
GS 04L53B00-01EN
June. 14, 2017-00
Safety and EMC Standards
• CSA:
CAN/CSA22.2 No.61010-1, installation category
II
*1
, pollution degree 2
*2
• UL:
UL61010-1 (CSA NRTL/C)
UL61010-2-201(CSA NRTL/C)
*3
• CE/EMCdirective:
EN61326-1compliance,ClassATable2
EN61000-3-2compliance
EN61000-3-3compliance
EN55011 Class A Group 1
• CE/Lowvoltagedirective
EN61010-1
Installation category II
*1
Pollution degree 2
*2
EN61010-2-201
*3
• EMCRegulatoryArrangementinAustraliaand
NewZealand(RCM):EN55011compliance,
Class A Group 1
• KCmarking:Electromagneticwaveinterference
preventionstandard,electromagneticwave
protectionstandardcompliance
*1 Installationcategory(overvoltagecategory)II:
Describesanumberwhichdenesatransient
overvoltagecondition.
Impliestheregulationforimpulsewithstand
voltage.
“II”appliestoelectricalequipmentwhich
issuppliedfromthexedinstallationlikea
distribution board.
*2 Pollutiondegree2:
Describesthedegreetowhichasolid,liquid,
or gas which deteriorates dielectric strength or
surfaceresistivityisadhering.
“2”appliestonormalindooratmosphere.
Normally,onlynon-conductivepollutionoccurs.
*3 Thisproductisdesignedasopenequipment
undertherelevantstandard,installitasfollows:
•InstalltheGX60/GMunitinapanelwithadoor.
•Theinstrumentationpanelorpanelusedfor
supportmustcomplywithCSA/UL/EN61010-
2-201ormustbeatleastIP1X(degreesof
protection)andatleastIK09.
• WEEEDirective:Compliant
Construction
• Frontpanel(terminal):Wateranddust-proof,
ComplieswithIEC529-IP20
• Material:Polycarbonate
• Color;
Front:Charcoalgreylight(Munsell10B3.6/0.3
equivalent)
Bezel:Smokeblue(Munsell4.1PB6.0/4.5
equivalent)
• Dimensions:45.2mm(W)x111mm(H)x107.1
mm(D)(D:includingterminalcover)
• Weight:Approx.0.18kg
Power Supply
SuppyfromGX/GP,GX60expandableI/O,or
GM90PSpowersupplymodule.
• Powerconsumption:1.8Worless
External Dimensions
Unit: mm (approx. inch)
45.2(1.78)
107.1
3(0.12)
8(0.31)
100(3.94)
31.9
82.1(3.23)
Normal Operating Conditions
Fornormaloperatingconditionsofthismodule,
pleaserefertotheGeneralSpecicationsofthe
device(GX/GP,I/OBaseUnit,orGM)thatthis
moduleismounted.
GXSpecications:GS04L51B01-01EN
GPSpecications;GS04L52B01-01EN
I/OBaseUnit(ExpandableI/O):ThisGeneral
Specications
GMSpecications:GS04L55B01-01EN
Transport and Storage Conditions
• Ambienttemperature:–25to70°C
• Ambienthumidity:5to95%RH(no
condensation)
• Vibration:10to60Hz,4.9m/s
2
maximum
• Shock:392m/s
2
maximum(inpackaged
condition)
Effects of Operating Conditions
None

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA GX20 and is the answer not in the manual?

YOKOGAWA GX20 Specifications

General IconGeneral
BrandYOKOGAWA
ModelGX20
CategoryRecording Equipment
LanguageEnglish

Related product manuals