GS 04L53B01-31EN
6
All Rights Reserved. Copyright © 2017, Yokogawa Electric Corporation 2017.06.26-00
*4 Thisproductisdesignedasopenequipmentunder
therelevantstandard,installitasfollows:
•
TheGX10/GX20isdesignedtobeinstalledinan
instrumentationpanel.Installitinalocationwhere
peoplecannottouchtheterminalscarelessly.
• TomaketheGP10/GP20complywiththe
relevantstandard,supportthepartsofthe
deviceotherthanthefront-panelcontrolarea
withaninstrumentationpanelorthelike,and
installitinalocationwherepeoplecannottouch
theterminalscarelesslyorinapanel.
• InstalltheGX60/GMunitinapanelwithadoor.
• Theinstrumentationpanelorpanelusedfor
supportmustcomplywithCSA/UL/EN61010-
2-201ormustbeatleastIP1X(degreesof
protection) and at least IK09.
• Tocomplywiththerelevantstandard,makesure
thatthestylenumbersoftheGX/GPmainunit,
GM90PS power supply module, and GX60 I/O
baseunitareatleast2.
• WEEE Directive
Compliant
Construction
• Front panel (terminal) area: Dust-proofanddrip-proof,
IEC529-IP20
•Material:Polycarbonate
• Color:
 Frontpanel:Lightcharcoalgray(Munsell10B3.6/0.3
equivalent)
 Case:Smokegray(Munsell4.1PB6.0/4.5equivalent)
•
Externaldimensions:45.2(W)x111(H)x133.1(D)mm
 (D:depthincludingtheterminalcover)
• Terminal type: M3 screw terminal
•Weight:Approx.0.3kg
Powersupply
SuppliedfromGX/GP,GX60,GM90PS
• Power consumption: 2.8 W or less
Isolation
Functional isolation
Analog input CH1
Analog input CH2
Analog output CH1
Analog output CH2
Digital input CH1 to 8
Digital output CH1 to 8
Internal
circuit
Terminalarrangement:M3screwterminal
No. Symbol No. Symbol No. Symbol
301 DI3 201 DI2 101 DI1
302 DI6 202 DI5 102 DI4
303 DI-COM 203 DI8 103 DI7
304 DO3 204 DO2 104 DO1
305 DO6 205 DO5 105 DO4
306 DO-COM 206 DO8 106 DO7
307 AI1( /A) 207 AI1(-/b) 107 AI1(+/B)
308 AI2( /A) 208 AI2(-/b) 108 AI2(+/B)
309 N.C. 209 AO1(-) 109 AO1(+)
310 N.C. 210 AO2(-) 110 AO2(+)
A/DCalibrationValue
YoucansavetwotypesofADcalibrationvalues:
factorydefaultvaluesanduser-denedvalues.Ifthere
isaproblemwithuser-denevalues,theycanbereset
tofactorydefaultcalibrationvalues.
ExternalDimensions
107.126
92
100
38
43.8
45.2
NormalOperatingConditions
•Ambienttemperature:–20to50°C
 However,useitwithintheambienttemperaturerange
denedfortheGX/GP/GMmainunit.
 Forthenormaloperatingconditionsofthismodule,
seethespecicationsoftheinstrument(GX/GP,
expandableI/O,GM)thatthismoduleisinstalledin.
GXspecications:GS04L51B01-01EN
GPspecications:GS04L52B01-01EN
ExpandableI/Ospecications:GS04L53B00-01EN
GMspecications:GS04L55B01-01EN
TransportandStorageConditions
•Ambienttemperature:-25to70°C
•Ambienthumidity:5to95%RH(nocondensation)
•Vibration:10to60Hz,4.9m/s
2
or less
•Shock:392m/s
2
maximum(inpackagedcondition)
EffectsofOperatingConditions
•
Effectsofambienttemperature:
 Fluctuationper10°Cchange
 ±(0.05%ofrdg+0.05%ofrange)orless
However,forKpvsAu7Fe,PR20-40:±(0.05%ofrdg+
0.1%ofrange)orless
 Cu10Ωorless:±(0.2%ofrange+0.1°C)orless
•Effectsofpoweructuation:
 Meetstheaccuracyspecicationsintheratedpower
supply range
•Effectsofmagneticeld:
 FluctuationinresponsetoamagneticeldofAC
(50/60Hz)400A/mis±(0.1%ofrdg+0.1%ofrange)
or less.