2
All Rights Reserved. Copyright © 2014, Yokogawa Electric Corporation
GS 04L53B01-01EN Nov. 10, 2015-00
INPUT/OUTPUT MODULE SPECIFICATIONS
ANALOG INPUT MODULE
(Model GX90XA or GX/GP main unit options /Uxx0)
• Numberofinputs:10
• Input Type:
Sufx ode Input Type Description
-U2 DCvoltage,standardsignal,thermocouple(TC),resistance
temperaturedetector(RTD),DI(voltage,contact),andDC
current (by adding an external shunt resistor)
Universal
-C1 DCcurrent(mA),DCcurrentstandardsignal(4-20mA) Current input
-L1 DCvoltage,standardsignal,thermocouple(TC),DI(volt-
age, contact), and DC current (by adding an external shunt
resistor)
Low withstand voltage
relay
-T1 DCvoltage,standardsignal,thermocouple(TC),DI(volt-
age, contact), and DC current (by adding an external shunt
resistor)
Electromagneticrelay
• Measurementinterval:100
*1*2
, 200
*1*2
,500ms
*1
, 1, 2, 5 s
• Inputrange:-5%ormoreand105%orless(accuracyisguaranteedintherangefrom0%to100%inclusive)
• Measurementrangesandaccuracies
*3
(However,thenumberofdisplaydigitscanbeincreasedbyscaling.)
*1
Cannotbespeciedfortheelectromagneticrelayscannertype(TypeSufxCode:-T1).
*2
Cannotbespeciedfor
L-modelDCV/TC/DI,scannertype
(TypeSufxCode:-L1).
*3
Thefollowingspecicationsapplytooperationoftherecorderunderstandardoperationconditions.
Temperature:23±2°C,Humidity:55%±10%RH,Powersupplyvoltage:90to132or180to264VAC,Powersupply
frequency:50/60Hz±1%,Warm-uptime:Atleast30min.Otherambientconditionssuchasvibrationshouldnotadversely
affect recorder operation.
Input Type Range Measurement range
Measurement accuracy (digital display)
Max.
resolution
of digital
display
A/D integration time: 16.7ms or more
*23
A/D integration time: 1.67ms
*24
DCV 20mV -20.000 to 20.000mV ±(0.05%ofrdg+12μV) ±(0.1%ofrdg+40μV) 1μV
60mV -60.00 to 60.00mV ±(0.05%ofrdg+0.03mV) ±(0.1%ofrdg+0.15mV) 10μV
200mV -200.00 to 200.00mV ±(0.05%ofrdg+0.03mV) ±(0.1%ofrdg+0.4mV) 10μV
1V -1.0000 to 1.0000V ±(0.05%ofrdg+1.2mV) ±(0.1%ofrdg+4mV) 100μV
2V -2.0000 to 2.0000V ±(0.05%ofrdg+1.2mV) ±(0.1%ofrdg+4mV) 100μV
6V -6.000 to 6.000V ±(0.05%ofrdg+3mV) ±(0.1%ofrdg+15mV) 1mV
20V -20.000 to 20.000V ±(0.05%ofrdg+3mV) ±(0.1%ofrdg+40mV) 1mV
50V -50.00 to 50.00V ±(0.05%ofrdg+0.03V) ±(0.1%ofrdg+0.15V) 10mV
Standard
signal
0.4-2V 0.3200 to 2.0800V ±(0.05%ofrdg+1.2mV) ±(0.1%ofrdg+4mV) 100μV
1-5V 0.800 to 5.200V ±(0.05%ofrdg+3mV) ±(0.1%ofrdg+15mV) 1mV
DC current 0-20mA 0.000 to 20.000mA ±(0.3%ofrdg+5μA) ±(0.3%ofrdg+90μA) 1μV
DC current
(standard
signal
4-20mA 3.200 to 20.800mA
TC
(Excluding
RJCac-
curacy)
R
*4
0.0 to 1760.0 °C ±(0.15 % of rdg + 1.0°C)
However, R, S; 0.0 to 800.0°C: ±2.2°C,
B; 400.0 to 800.0°C: ±3.0°C
Accuracy at less than 400.0°C is not guaranteed.
±(0.2 % of rdg + 6.0°C)
However, R, S; 0.0 to 800.0°C: ±7.6°C,
B; 400.0 to 800.0°C: ±11.0°C
Accuracy at less than 400.0°C is not guaranteed.
0.1°C
S
*4
0.0 to 1760.0 °C
B
*4
0.0 to 1820.0 °C
K
*4
-270.0 to 1370.0 °C ±(0.15 % of rdg + 0.7°C)
However, -200.0 to 0.0°C: ±(0.35 % of rdg + 0.7°C)
Accuracy at less than -200.0°C is not guaranteed
±(0.2 % of rdg + 5.0°C)
However, -200.0 to 0.0°C: ±(3 % of rdg + 5.0°C)
Accuracy at less than -200.0°C is not guaranteed
0.1°C
-200.0 to 500.0 °C
E
*4
-270.0 to 800.0 °C ±(0.15 % of rdg + 0.5°C)
However, -200.0 to 0.0°C: ±(0.35 % of rdg + 0.5°C)
Accuracy at less than -200.0°C is not guaranteed
±(0.2 % of rdg + 4.0°C)
However, -200.0 to 0.0°C: ±(2 % of rdg + 4.0°C)
Accuracy at less than -200.0°C is not guaranteed
0.1°C
J
*4
-200.0 to 1100.0 °C
T
*4
-270.0 to 400.0 °C ±(0.15 % of rdg + 0.5°C)
However, -200.0 to 0.0°C: ±(0.35 % of rdg + 0.5°C)
Accuracy at less than -200.0°C is not guaranteed
±(0.2 % of rdg + 2.5°C)
However, -200.0 to 0.0°C: ±(2 % of rdg + 2.5°C)
Accuracy at less than -200.0°C is not guaranteed
0.1°C
N
*4
-270.0 to 1300.0 °C ±(0.15 % of rdg + 0.7°C)
However, -200.0 to 0.0°C: ±(0.7 % of rdg + 0.7°C)
Accuracy at less than -200.0°C is not guaranteed
±(0.3 % of rdg + 6.0°C)
However, -200.0 to 0.0°C: ±(5 % of rdg + 6.0°C)
Accuracy at less than -200.0°C is not guaranteed
0.1°C
W
*5
0.0 to 2315.0 °C ±(0.15 % of rdg + 1.5°C) ±(0.3 % of rdg + 14.0°C)
However,morethan1000.0°C:±(0.8%ofrdg
+ 9.0°C)
0.1°C
L
*6
-200.0 to 900.0 °C ±(0.15 % of rdg + 0.5°C)
Less than 0.0°C: ±(0.5 % of rdg + 0.5°C)
±(0.2 % of rdg + 4.0°C)
Less than 0.0°C: ±(3 % of rdg + 4.0°C)
0.1°C
U
*6
-200.0 to 400.0 °C ±(0.15 % of rdg + 0.5°C)
Less than 0.0°C: ±(0.7 % of rdg + 0.5°C)
±(0.2 % of rdg + 2.5°C)
Less than 0.0°C: ±(3 % of rdg + 2.5°C)
0.1°C
WRe3-25
*7
0.0 to 2320.0 °C ±(0.2 % of rdg + 2.5°C) ±18.0°C
More than 2000.0°C: ±0.9 % of rdg
0.1°C
KpvsAu7Fe
*8
0.0 to 300.0 K ±(0.15 % of rdg + 2.0 K) ±(0.2 % of rdg + 7.0 K) 0.1 K
PLATINEL II
*8
0.0 to 1395.0 °C ±(0.25 % of rdg + 2.3°C) ±(0.25% of rdg + 8.0°C) 0.1°C
GX90XA