EasyManua.ls Logo

Zeiss Axiotech - Page 75

Zeiss Axiotech
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Axiotech and Axiotech
vario
Carl Zeiss
B 40-020 e 06/99 3-9
Focus on the specimen via the coaxial drive (3-7/4). If possible, always move the specimen away from
the objective to avoid a collision between the objective and the specimen.
Move aperture diaphragm (3-7/2) in center position (approx. half opened or half closed) by pulling out
the pushrod.
Pull out the pushrod or adjust the setting wheel to set (reduce) the diameter of the luminous-field
diaphragm (3-7/1) until it is visible in the field of view (3-8/A).
Use the coaxial focusing drive (3-7/4) to refocus on the edge of the luminous-field diaphragm (3-8/B)
and use screws (3-7/3) to center the luminous-field diaphragm in the edge of the field of view (3-8/C).
Then open the luminous-field diaphragm (3-7/1) until it just disappears behind the edge of the field of
view (3-8/D).
To set the aperture diaphragm (image contrast), remove one eyepiece from the tube and look into
the tube with your naked eye, or insert the centering telescope instead of the eyepiece.
Center aperture diaphragm using centering screws ( 3-1/7), (3-2/7) or (3-3/7) and set pushrod (3-1/6),
(3-2/6), (3-3/9) or (3-7/2) to approx. 2/3 of the exit pupil diameter of the objective for specimens of
medium contrast ( 3-8/E). In most applications, this setting of the aperture diaphragm provides
optimum contrast at almost full resolution and is therefore the best compromise for the human eye.
Finally, refocus on the specimen via the coaxial coarse and fine drive ( 3-7/4) and match the image
brightness to the reflected-light specimen.
Since field size and objective aperture change after every objective change, the settings of the
luminous-field diaphragm and the aperture diaphragm described above must be repeated.
Do not use the aperture diaphragm for adjustment of the image brightness, but use
illuminance control (3-4/3) or the attenuation filter instead, as described under 2.9.1!
Fig. 3-8 Setting of diaphragm images according to KÖHLER

Table of Contents

Related product manuals