Axiovert 40
B 46-0031 e 11/03 2Ć3
(3) Inserting diaphragm sliders
• Remove dust protection slider (2Ć4ā/ā1) from the
slider mount and replace it by the
- Slider Ph, H, Ph (2Ć4ā/ā2) or the
- Slider Ph, H, Var (2Ć4ā/ā3) or the
- Slider Var, H, Var (not illustrated) or the
- Slider Ph, H, PlasDIC (2Ć4ā/ā4).
Push it into the middle position (brightfield).
NOTE The label on the slider must be legible
upright and the inserted diaphragms
must correspond to the condenser/
objective combination used (refer to
the table on page 2Ć14).
(4) Mounting the HBO 50 illuminator
• Attach the HBO 50 illuminator (2Ć5ā/ā3) equipped
with mercury vapor short-arc lamp to the correĆ
sponding dovetail mount of the Axiovert 40 CFL
and tighten the fastening screw by means of the
SW 3 ball-headed screwdriver.
For information on lamp replacement, refer to
Section 3.2 (3).
• Establish the connection between the illuminator
and the HBO 50 power supply (2Ć5ā/ā1). Connect
the power cable (2Ć5ā/ā2) to the power outlet.
• Switch on the power supply.
Fig. 2Ć4 Inserting diaphragm sliders
1
2
3
4
Fig. 2Ć5 Mounting the HBO 50 illuminator