EasyManua.ls Logo

Zeiss Axiovert 40 - Incident-Light Fluorescence; Observation in Incident-Light Fluorescence

Zeiss Axiovert 40
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Axiovert 40
B 46-0031 e 11/03 2Ć19
2.4.6 Incident-light fluorescence
Incident-light fluorescence requires the use of the Axiovert 40 CFL microscope with the following compoĆ
nents:
FL incident-light equipment, HBO 50 illuminator with mbq ac-z transformer, 3-position reflector slider P&C
and reflector module FL P&C (equipped with set of filters for fluorescence excitation).
Select the specimen area to be examined in transmitted light brightfield or phase contrast. To do this, push
reflector slider (2Ć24ā/ā10) to the blank position, switch on halogen bulb (2Ć24ā/ā2) and move diaphragm slider
(2Ć24ā/ā3) to brightfield or phase contrast position.
Switch on the HBO 50 lamp (2Ć24ā/ā5) at the power supply (2Ć24ā/ā4), but block the light path using the addiĆ
tional filter slider (2Ć24ā/ā6), which in its mid-position does not allow the light to pass.
After you have selected the specimen area of interest, switch off the halogen bulb.
With the 3-position reflector slider P&C (2Ć24ā/ā10) move the required filter combination (2Ć24ā/ā9) into the
light path and release the light path by removing the additional filter slider (2Ć24ā/ā6).
Move the push-pull rod (2Ć24ā/ā7) to close the luminous-field diaphragm until it is visible in the field of view.
Center it by means of centering screws (2Ć24ā/ā8) and open it up to the edge of the field of view.
In the additional filter slider (2Ć24ā/ā6) of the FL incident-light equipment, you can insert additional 25-mm
excitation filters, which must be held by means of O-rings.
Fig. 2Ć24 Observation in incident-light fluorescence

Table of Contents

Related product manuals