EasyManua.ls Logo

Zeiss Axiovert 40 - Connection to the Power Outlet

Zeiss Axiovert 40
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Axiovert 40
B 46-0031 e 11/03 2Ć5
(6) Connection to the power outlet
CAUTION Verify the operating voltage indicated on the rear panel of the instrument agrees with the
available line voltage.
Plug power cable (2Ć7ā/ā4) into the instrument socket and connect to the power outlet.
Switch on the instrument at the power switch (2Ć7ā/ā3) on the right of the instrument.
NOTE When the instrument is switched off, the "0" marking is visible on the power switch.
The instrument is ready for operation when the microscope lamp is on.
In the event of a defect, check the two fuses (2Ć7ā/ā2):
for 100240 V AC: T 1 A/H, 5 x 20 mm (acc. to IEC 127)
For this purpose, remove the fuse holder (2Ć7ā/ā1) from the housing by pressing the two springs simultaneously
in the direction indicated by the arrows.
CAUTION Only use the correct fuses.
Fig. 2Ć7 Connection to the power outlet
1
2
3
4

Table of Contents

Related product manuals