PhySys Edition SD Service Manual
Rev. B
20.08.2013
Page 47/52
Service Manual PhySysSD_RevB_20.08.2013.do
c
Error /Error
code(xxxxx)
Description
Error /Error
code(xxxxx)
Description
-91 FLASH checking error 100 RAM test failed
-92 runtime RAMCODE checking error 101 a critical variable was changed unexpectedly
-93 IRQ timing error 102 an EM lookup table is corrupted
-94 Ticker tasks not executed properly 103 bootloader: there is no Application in FLASH
-95 Idle Task 104 a supervised variable's value is out of range
-96 watchdog test failed 105
105 the firmware is running on the wrong
hardware (EM/US)
-97
there was an unexpected reset by external
watchdog
-98
there was an unexpected reset by internal
watchdog
10.4.4. Cable Control Problems
When increasing the current to the patient, the message “Cable control” appears.
Remedy:
¾ Check, if the electrode cables are correctly connected
¾ Inspect the electrode cables, electrode clamps
¾ Replace electrode cables if they are contaminated or corroded
¾ Inspect the suction-wave cables