PhySys Edition SD Service Manual
Rev. B
20.08.2013
Page 52/52
Service Manual PhySysSD_RevB_20.08.2013.do
c
12.2. VacoP device
VacoP serial number: ………………………………….
Serial number of board: ………………………...………..
Visual inspection:
1x Fuse (0,4AT fuse at 100-120V device; 0,2AT fuse 230-240V device)
O
ok
1x Fuse (1,25AT) on the mainboard
O
ok
230-240V device, jumpers JP1001, JP1002 are set.
O
ok
100-120V device, jumper JP1003, JP1004 are set
O
ok
Internal wiring
O
ok
All screws present and tight
O
ok
All silicon tubes firm attached
O
ok
Housing (no damage, labels legible, no dirt)
O
ok
Electrical safety:
Ground leakage current (max.500PA)
O
ok
Ground leakage current 1st error (max.1000PA)
O
ok
Housing leakage current (max.100PA)
O
ok
Housing leakage current 1st error (max.500PA)
O
ok
Patient leakage current (channel 1 max.100PA)
O
ok
Patient leakage current 1st error (channel 1 max.500PA)
O
ok
Patient leakage current (channel 2 max.100PA)
O
ok
Patient leakage current 1st error (channel 2 max.500PA)
O
ok
Function test:
Communication VacoP with PhySys Edition SD ( Vacuum Panel on the therapy window)
O
ok
Device test:
Stimulation current channel1 with suction electrodes (GA01,CV)
O
ok
Stimulation current channel2 with suction electrodes (GA01,CV)
O
ok
Vacuum test/ tightness test:
(Therapy vacuum >40kPa, pump start)Æ No repeated pressure regulation
O
ok
Pulsation
O
ok
Test run with different parameter’s settings:
O
ok
All items checked
O
ok
Date:
______________________________
Name:
_______________________________
Signature:
____________________________
Please observe the legal regulations for repairs and maintenance of medical devices in your country.