PhySys Edition SD Service Manual
Rev. B
20.08.2013
Page 49/52
Service Manual PhySysSD_RevB_20.08.2013.do
c
10.6. Test Run
10.6.1. PhySys Edition SD
The following is required after each PhySys Edition SD repair (module replacement):
¾ The device must be visually inspected (loose parts, plug connector, etc.)
¾
A safety check must be performed (see chapter 12.2).
10.6.2. VacoP
The following is required after each VacoP repair (module replacement):
¾ The device must be visually inspected (loose parts, cable fixation, plug connector, silicone tube,
Jumper, fuses, etc.)
¾ A test run with a therapy time of 15 minutes must be carried out.
¾
A safety check must be performed (see chapter 12.2).