10.4. Error Messages
Important hardware components are tested while the device is starting up and at times while the device is operating.
If an error is detected during the test, on-going treatment sessions are terminated, all inputs and outputs are disabled
and an error message appears. These error measures protect the patient by bringing the device into a safe status.
¾ The error messages appear on the display. Therapy is no longer possible; there is only the
option to switch off the device.
¾ Many errors can be resolved by shutting off the device, waiting 1 minute and re-starting the
device. These errors are referred to as fleeting errors, which do not influence continued
operation of the device.
¾ If error messages indicating excessive temperature appear, the device should be switched
back on only after 30 minutes have passed, since an excessive temperature situation exists.
The PhySys Edition SD requires time to cool down. This may be caused by high outside
temperatures.
10.4.1. General description
Error xxxxx/y/zz
Xxxx Up to 5digits
- numbers from 1-999 indicate errors of the Slave processors
- Numbers starting at 1000 and higher indicate errors of the Master Processor
y Shows which processor messages the error
zz More detailed error description (Zimmer internal use only)
10.4.2. Master Processor
Error /Error
code(xxxxx)
Description
Error /Error
code(xxxxx)
Description
-1001 general error -1014 An unexpected watchdog reset occurred
-1002 Cancel the Sreen or Popup -1015 Stack boundary violation
-1003 illegal case state in graph -1016 an exception has occurred
-1004 illegal action state in graph -1017 A timeout has occurred
-1005 Value was not in range -1018 illegal date
-1006 Nullptr detected -1019 illegal time
-1007 not enough heap memory -1020 No button of This window was pressed
-1008 unknown Error -1021 Window which should be created still exists
-1009 Function called in illegal context -1022 Channel used here is not the right one
-1010 execution flag was not false, smthg wrong -1023 No Valid Value is passed
-1011 Ticker tasks overflow -1024 Unknown id received from IrDA
-1012 Slow Ticker tasks overflow -1025
A program has been loaded in the Proghandler
before trying again to load a program
-1013 Watchdog test failed, did not reset the CPU -1026
No program has been loaded in the Proghandler
before trying to access the program
-1027 error in GUI memory management -1052 improper command sequence
-1028 No effect has been found -1053 low programming voltage
-1030 no ACK received on I2C bus -1054 block erase error
-1031 EERPOM write timeout -1055 clear block lock-bits error
-1032 EERPOM verify error -1056 set block lock-bits error
-1035 no configuration data in EEPROM -1057 write error
-1040 Transmit is not ready yet -1058 erase, prog timeout
-1041 Receiving Timeout -1059 verify failure