PhySys Edition SD Service Manual
Rev. B
20.08.2013
Page 46/52
Service Manual PhySysSD_RevB_20.08.2013.do
c
10.4.3. Slave Processor
Error /Error
code(xxxxx)
Description
Error /Error
code(xxxxx)
Description
0 no error -38 selftest: 200V supply too high
-1 general error -39 selftest: 12V supply too low
-2 illegal case state in graph -40 selftest: 12V supply too high
-3 illegal action state in graph -41 selftest: 12VA supply too low
-4 attempt to deref. a NULL pointer -42 selftest: 12VA supply too high
-10
command was not executed because system
is locked -50 selftest: ImVolt not 0 though switched OFF
-11 unknown command received -51
selftest: Head Overvoltage detected on selftest --
> head probably not connected
-12
communication request for non-implemented
parameter -52 selftest: Im-voltage lower than expected
-13 attempt to set a value that is not allowed -53 selftest: Im-voltage higher than expected
-14 attempt to set an idxed value with a wrong idx -54
US too high head voltage was measured -->
head probably disconnected
-15 attempt to read -55
US too low head voltage was measured -->
probably IVolt failure
-16
cannot execute CMD, still busy with previous,
please repeat -56
US time between measurement pulses was
much longer than allowed
-17 attempt to write to a read-only parameter -57 US communication with US-Head failed
-18
illegal or incomplete EM parameter set ->
cannot start therapy -58 US head overtemperature
-19
illegal or incomplete US parameter set ->
cannot start therapy -60 EM overcurrent detected by SW in fast context
-20 selftest: 200V supply too low -61 EM overcurrent detected by HW
-21 selftest: 1.5V supply -62 EM CC: high Zload, cable break
-22 selftest: ImVolt not 0 though switched OFF -63 EM CV: high Zload
-23 selftest: ImCurr not 0 though switched OFF -64 EM CV: low Zload, short circuit
-24 selftest: ImVolt lower than expected -65 EM CC: low Zload, short circuit
-25 selftest: ImVolt higher than expected -66 EM overcurrent detected by SW
-26 selftest: Reference resistor A voltage is not 0 -67 EM output voltage limit reached
-27 selftest: Reference resistor B voltage is not 0 -70 US-Head EEPROM: unexpected NACK
-28
selftest: Reference resistor voltage is lower
than expected -71 US-Head EEPROM: verify error ofter writing
-29
selftest: Reference resistor voltage is higher
than expected -72 US-Head EEPROM: write timeout
-30 selftest: EM current is not 0 -73 US-Head EEPROM: CRC error
-31 selftest: EM current is lower than expected -75
US-Head temperature sensor: unexpected
NACK
-32 selftest: EM current is higher than expected -76
US-Head temperature sensor: illegal value
received
-33 selftest: EM voltage A is not 0 -80 H-bridge over temperature
-34 selftest: EM voltage B is not 0 -81 IM overcurrent
-35 selftest: failed to set relays-enable-flipflop -82 PC_12V error
-36 selftest: failed to reset relays-enable-flipflop -83 PC_12V_A error
-37 selftest: hardware overcurrent detector failed -84
Warning to user: the DSP software is a debug
version
-90 runtime stack checking error -99 no data received for too long a time