20. Specifications
BM-20 BM-22 BM-252 BM-500 BM-300 BM-200
22 g
Weighing capacity 22 g
5.1 g
250 g 520 g 320 g 220 g
22.00008 g
Maximum display 22.000084 g
5.100009 g
250.00084 g 520.0084 g 320.0084 g 220.0084 g
0.01 mg
Minimum display 0.001 mg
0.001 mg
0.01 mg 0.1 mg
0.01 mg
Repeatability
(Standard deviation)
0.0025 mg
0.004 mg
0.03 mg 0.2 mg 0.1 mg
±0.02 mg
Linearity ±0.010 mg
±0.010 mg
±0.10 mg ±0.5 mg ±0.3 mg ±0.2 mg
Approx. 8 seconds
Stabilization time
(Typical at FAST)
Approx. 10 seconds
Approx. 3.5 seconds
Sensitivity drift ±2 ppm/°C (10 °C to 30 °C, Automatic self calibration: OFF)
Operating environment 5 °C to 40 °C, 85 % RH or less (No condensation)
Display refresh rate 5 times/second or 10 times/second
Environmental sensors
Resolution: Temperature ±1.5 °C, Humidity ±10 %, Barometric pressure ±10 hPa
Applicable range: 5 °C to 40 °C
Minimum unit
mass
0.1 mg
Counting
mode
Number of
samples
10, 25, 50 or 100 pieces
Minimum 100%
reference mass
1 mg 10 mg
Percent
mode
Minimum display 0.01 %, 0.1 %, 1 % (Depends on the reference mass stored.)
Interface USB and RS-232C
External calibration
weight
20 g
10 g
5 g
2 g
1 g
20 g
10 g
5 g
2 g
1 g
200 g
100 g
50 g
20 g
10 g
500 g
300 g
200 g
100 g
50 g
300 g
200 g
100 g
50 g
200 g
100 g
50 g
Size of loading area for
weighing mm
Weighing pan:25,
Filter weighing: 50, 95
Micro tube holde
: 12
Weighing pan:90
Micro tube holder: 12 (Option)
External dimensions 259(W) x 466(D) x 326(H) mm
Power supply &
AC adapter type
Power consumption: Approx. 30 VA (supplied to the AC adapter )
Confirm that the adapter type is correct for the local voltage and power
receptacle type.
Mass Approx. 10 kg
The operating environment with no excessive changes in ambient temperature and humidity, no
vibration and drafts, no influence by magnetic fields and static electricity.
The internal mass may change due to corrosion or other damage caused by the operating environment,
or due to aging. Check the internal mass periodically and correct the internal mass value if necessary.
*1: Even when a tare weight beyond the precision range is placed, the weighing sample can be weighed in
the precision range if using the smart range.
*2: Repeatability using a 1 gram weight within the weighing capacity.
*3: Repeatability using a 100 gram weight within the weighing capacity.