9.2. Initializing the Balance
This function restores the following parameters to the factory settings.
Calibration data
Function table (excluding the time and date)
Sample unit mass value (counting mode),
100% reference mass value (percent mode)
The data that is stored in the balance using the data memory function
External weight value
Function switch settings ("9.1. Permit or Inhibit")
Liquid density and temperature in the density mode
Note Be sure to calibrate the balance after initialization.
1 Press the ON:OFF key to turn the display off.
2 While pressing and holding the PRINT key and the
RANGE key, press the ON:OFF key to display p5 .
3 Press the RANGE key to display Clr .
4 Press the PRINT key to display Clr no .
To cancel the operation, press the CAL key.
5 Press the RE-ZERO key to display Clr go .
6 Press the PRINT key to initialize the balance.
The balance will automatically return to the weighing mode.