8.6. Calibration Test Using an External Weight
This function checks the accuracy of weighing using an external weight.
This is available only when the "GLP output (info)" parameter is set to "1", "2" or "3".
Calibration test does not perform adjustment.
1 Connect the AC adapter and warm up the balance for at least
one hour with nothing on the weighing pan.
2 Press and hold the CAL key until CC out is displayed,
then release the key.
3 When CC 0 is displayed, proceed as follows:
If you want to change the target weight, press the RANGE
key and proceed to Step 4. A list of usable weights is on
page 36.
If the current target weight value is used, proceed to Step 5.
4 Specify the target weight value as follows:
RANGE key
........
To switch blinking digits.
RE-ZERO (+)key
...
To select the target weight or adjust the
weight value. Refer to page 36.
PRINT key
..........
To store the new target weight value.
Even if the AC adapter is removed, the
data is maintained in non-volatile memory.
CAL key
.............
The key to cancel the operation and
return to CC 0 .
5 Confirm that there is nothing on the pan and press the
PRINT key. The balance measures the zero point.
Do not allow vibration or drafts to affect the balance.
The balance displays the target weight value.
6 Place the displayed target weight on the pan and press
the PRINT key. The balance measures the target weight.
Do not allow vibration or drafts to affect the balance.
7 end is displayed.
Remove the weight from the pan.
8 glp is displayed and the calibration test report is output and
stored. Refer to "GLP output (info)" and "Data memory
(data)" of the function table in "11.2. GLP Report" and "12.
Data Memory".
9 The balance will automatically return to the weighing mode.
MODE (-)ke