32 BM series
6.6. Monitoring the Clock and Environmental Sensors
In the weighing mode, the built-in clock and environmental sensors (temperature, humidity,
barometric pressure) can be monitored and temperature data can be output in the function
settings.
Specifications of the environmental sensors
Sensor Resolution Measurement range Applicable range
Temperature ±1.5 °C 5 °C to 40 °C
Humidity ±10 % 0 % to 100 %
Barometric pressure ±10 hPa 300 hPa to 1100 hPa
5 °C to 40 °C
Notes
The environmental sensors are built into the balance casing. Therefore, these values are
not in accordance with the environmental condition.
Refer to "20. Specifications" for the balance operating conditions.
Operation
Press the SELECT key in the weighing
mode. The current time, date and sensor
values (temperature, humidity, barometric
pressure) are displayed for several seconds
in order. The balance returns to the
weighing mode automatically.
When pressing the SELECT key
while the current data is displayed,
the display proceeds to the next
data.
When pressing the CAL key
while the current data is displayed,
the balance returns to the weighing
mode.
Notes
When adjusting the built-in clock, refer to "10.7. Clock and Calendar Function".
When adding the time and date to the output data, refer to “Time/Date output (5-td)” of
"10.2. Details of the Function Table".
When adding temperature data to the output data, refer to “Temperature output (5-tp)”
of "10.2. Details of the Function Table".
Humidit
y
Barometric pressure
T
e
m
pe
r
atu
r
e
Weighing mod
e
Tim
e
5 second
s
Dat
e
5 second
s
Environmental sensor
s
10 second
s
Weighing mod
e