BM series 49
Class Item and Parameter Description
off No limit
t-Up
Timeout
on 1 second
Selects the wait time to receive
a command.
off No output
5if
Serial
interface
erCd
AK, Error code
on Output
AK: ASCII code 06h
0 Water temperature
d5 fnc
Density
function
ldin
Liquid density input
1 Liquid density
Available only when density
mode is selected
Unit
Unit
Refer to "5. Weighing Units".
C5 in
Internal mass correction
Displayed only when the internal mass value correction
switch is set to 1. Refer to "8. Calibration ".
id
ID number setting
Refer to "11. ID Number and GLP Report".
0 3 seconds
1 10 seconds
ion
Neutralization time
2 Manual operation (max. 10 min.)
Refer to "6.5. Ionizer" ".
off The ION key is not used.
8922
AD8922A control
on Assigns the ION key to the MODE key on AD-8922A.
0 Assigns the PRINT key and RE-ZERO key.
1 Assigns the PRINT key and ION key.
ionfnc
Ionizer
function
e-5
External control input
2 Assigns the ION key and RE-ZERO key.
off Not used
door
Door sensor
function
d-dt
Door sensor
on Used
Refer to "6.1. Basic Operation"
0 Not used
1 Used (Limits weighing operations)
locfnc
Password
function
pa55
#4
Password function
2 Used (Allows basic weighing)
Refer to "15. Password
Function”.
ADMIN
Password input by the administrator
loc no.
Password
U5e
R
01
U5e
R
10
Password input by user 1
Password input by user 10
#5
:
Factory settings.
#4
Only “Not used (0)” and “Used (1) are available for balances with software version 1.50 to 1.64.
All the parameters are available for balances with software version 1.660 or later.
#5
Available for balances with software version 1.660 or later.
For balances with software version 1.50 to 1.64, only the password input by the administrator is
available.
10.3. Description of the Class "Environment, Display"
Condition ( Cond )
Cond 0 This parameter is for sensitive response to the fluctuation of a mass value. Used for
powder target weighing, weighing a very light sample or when quick response weighing
is required. After setting, the balance displays FAST .
Cond 2 This parameter is for stable weighing with slow response. Used to prevent a weight
value from drifting due to vibration or drafts. After setting, the balance displays SLOW .
Note In automatic response adjustment, this parameter is selected automatically.