Calibration report using an external weight
This is the GLP report when the balance is calibrated using the external weight.
Calibration test report using an external weight
This is the GLP report when checking the weighing accuracy of the balance with the external weight.
(Adjustment is not performed)
CR
LF
Setting of
info 1
PC format
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
<TERM>
Setting of
info 1
Printer format
A & D
MODEL GX-10002A
S/N 123456789
ID LAB-0123
DATE 2017/12/31
TIME 12:34:56
CALIBRATED(EXT.)
CAL.WEIGHT
+100000.00 g
REMARKS
SIGNATURE
A & D
MODEL GX-10002A
S/N 123456789
ID LAB-0123
DAT E 2017/12/31
TIME
CALIBRATED(EXT.)
CAL.WEIGHT
+10000.00 g
REMARKS
<TERM>
<TERM>
SIGNATURE
<TERM>
<TERM>
----------------
<TERM>
<TERM>
Space, ASCII 20h
Terminator, CR LF or CR*
Carriage return,ASCII 0Dh*
Line feed, ASCII 0Ah*
Calibration weight
Remarks
Signature
<TERM>
<TERM>
<TERM>
<TERM>
<TERM>
<TERM>
<TERM>
<TERM>
<TERM>
<TERM>
<TERM>
<TERM>
CR
LF
Space, ASCII 20h
Terminator, CR*LF or CR
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
<TERM>
A & D
MODEL GX-10002A
S/N 123456789
ID LAB-0123
DATE 2017/12/31
TIME 12:34:56
CAL.TEST(EXT.)
ACTUAL<TERM>
+0.00 g
+9999.95 g
TARGET
+100000.00 g
REMARKS
<TERM>
<TERM>
SIGNATURE
<TERM>
<TERM>
Manufacture
Model
Serial number
ID number
Date
Time
Calibration test
Zero point value
Target weight
value
Target weight
Remarks
Signature
Setting of
info 1
Printer format
Setting of
info 1
PC format
A &
D
MODEL GX-10002A
S/N 123456789
ID LAB-0123
DATE 2017/12/31
TIME 12:34:56
CAL.TEST(EXT.)
ACTUAL
0.00 g
+9999.95 g
TARGET
+10000.00 g
REMARKS
SIGNATURE