EasyManua.ls Logo

Abbott Prodigy 3855 - Page 32

Abbott Prodigy 3855
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
To Set Bolus Smulaon Mode
Bolus mode allows the physician to set a me period for therapy to be on and a me period for therapy to be
o. Aer a smulaon period expires, a lockout phase begins during which no smulaon can be acvated.
Aer the lockout, the paent can manually restart smulaon for the set me period. When a program is set
to bolus mode, the leer B is displayed on the Operaonal Display screen with informaon about the “on” or
ome remaining.
Table 5. Descripon of the Operaonal Display screens in bolus mode
Operaonal Display Screen Descripon
Smulaon is on for the me displayed (in hours and minutes or in minutes
and seconds).
Smulaon is in a lockout phase, where smulaon is o and cannot be
restarted unl the displayed me expires (in hours and minutes or in
minutes and seconds).
Smulaon is o, the lockout phase is nished, and the bolus program can
be restarted.
NOTE: When the bolus lockout me expires, the paent must use the
programmer to restart smulaon.
To set the smulaon mode to bolus:
1.
From the Program screen, press the Balance key unl it highlights the SmMode eld. The SmMode
1.
screen appears.
2.
Press the Amplitude Increase key unl BOLUS appears in the SmMode eld.
2.
Figure 66.
Select BOLUS in the SmMode eld
3. Press the Balance key to highlight the On eld.3.
Figure 67. Highlight the On eld
NOTE:
Bolus mode can provide up to 18 hours of “on” me and 18 hours of “ome.
Since the paent cannot start the bolus smulaon during the lockout phase, many clinicians enter a
one-second “ome so that the paent can restart smulaon when desired.
4. Press a Scroll key to highlight the desired me eld (H for hours, M for minutes, or S for seconds), and4.
press the Amplitude Increase key or Amplitude Decrease key to increase or decrease the value.
5.
Aer you nish seng all values for “on” me, press the Balance key to highlight the O eld.
5.
6. Follow step 4 to set the value for the “ome.6.
7. Press the Balance key to highlight the MagnetUse eld.7.
8.
Press the Amplitude Decrease key to change the magnet mode.
8.
28

Table of Contents