Page B'of (()XM_unity_Installation&Maintenance_Manual Ed.1.3c2013
B4;4
take several exposures and check that:
there are no errors
the led of the selected tube head is lit
the exposure button switch led is lit over the whole duration of the acoustic signal (beep)
B4=4
To check the power absorbed by the radiographic system an AC ammeter must be used
Connect the instrument to the power supply line
2. Set the xray to 65kV 7mA 2 s on the control panel
3. Carry out an exposure and read the current value on the instrument
/7
**
≤
≤≤
≤
B+LI@@#>G@3(
+)'(*5
B4@4
To check the electrical system an AC voltmeter must be used:
1. Measure the voltage from mains terminals
2. Measure and record the line voltage
3. Set an exposure of 65kV 7mA 2 s on the control panel
4. Proceed to the exposure and measured the line voltage during exposure
/7
13
@