Here below these instructions will explain how to perform Xray tube calibration procedure.
1. Press and keep pressed key mA for 5 seconds until message “TUBE CALIBRATION” appears on the display
2. Take a safety position far from the xray beam, press and keep pressed the exposure key . On the display will appear
“CAL. IN PROGRESS” message, Exp. Led is lit and an intermittent acoustic sound is emitted.
3. At the end of the procedure, acoustic sound stops and on the display appears “CAL. SUCCESS” message.
4. After waiting the pause time (it will be longer than the normal rest time), the timer back to standard functioning and it is
ready for a new exam.
+$'& *)*$
***