Configuration and Activation 4-9
dialog based on the Default Location (designated on the
Settings dialog), without changing the location name.
3. Select TRITON+ as the type of monitor the location from the
Series drop-down list. If you selected an existing site from the
Location Name drop-down list, this field will automatically
display the monitor series corresponding to that location.
4. Enter or edit the monitor serial number in the Serial Number
field.
5. Verify or select the interval at which you want the monitor to
take measurements and log data from the Sample Rate drop-
down list.
6. Verify or select the interval at which you want the monitor to
take measurements and log data when a threshold has been
crossed from the Fast Rate drop-down list.
7. Verify or select the interval at which you want the monitor to
scan the sensors to detect whether a threshold has been crossed
in the Scan Rate drop-down list.
8. Enter or edit the distance (in inches or mm) from the manhole
rim to the bottom center of the invert for the Manhole Depth.
9. Verify or select the method through which you will
communicate with the monitor from the Connect drop-down
list.
Serial Choose this option to perform direct, serial
communication with the monitor.
Note: Always use Serial communications to perform the
initial monitor activation or after a deactivation.
Wireless Choose this for TRITON+ monitors with SIMs
with static IP addresses for performing wireless
communication.
10. (applicable only to Wireless communications) Enter the IP
address corresponding to the monitor in the IP Address field.
The field designation displayed depends on the method of
communication selected in the Connect field.