4-32 ADS TRITON+
®
Manual
z Electronic Offset Enter the difference that exists between
the pressure depth sensor readings and the manually measured
depth. The default value is 0.
z Store Data Select this checkbox to ensure the monitor logs
pressure depth data to memory. This option is selected by
default.
z Automatic Calibration Select this checkbox to ensure the
monitor automatically calibrates questionable or erroneous
pressure depth data to read consistently with the ultrasonic
depth data. The monitor will maintain a running average of the
difference between the pressure depth and the ultrasonic depth
data throughout the day. At midnight each day, the monitor
applies a new offset based on the most recent difference in the
running average.