11-56 ADS TRITON+ Manual
Range from ultrasonic depth subsystem is significantly
greater than manually measured range.
Physical offset may be incorrect.
Pipe height may be incorrect.
Electronic offset may be incorrect.
Sensor may not be level.
Foam, extreme condensation, or other substance may
be absorbing the pulse.
Sensor may be faulty.
Verify the physical offset.
Verify electronic offset and adjust if necessary.
Make sure the sensor is level and in good
condition.
Check the hydraulic conditions in the pipe.
Fire the sensor at a shorter distance onto a hard surface
to confirm accuracy.
Contact your regional ADS representative.