Maintenance and Troubleshooting 11-57
Problem
Range from the ultrasonic depth subsystem is too short
(but not zero).
Possible
Causes
Pipe height may be incorrect.
Electronic offset may be incorrect.
Sensor may be dirty.
Sensor may be faulty.
Possible
Solutions
Verify the pipe height.
Verify electronic offset and adjust if necessary.
Clean the sensor.
Replace the sensor.
Contact your regional ADS representative.
Problem
Raw ultrasonic depth data shows depths greater than
the pipe diameter.
Possible
Causes
Physical offset may be incorrect.
Pipe height may be incorrect.
Electronic offset may be incorrect.
Pipe may be surcharged.
Sensor may be dirty.
Sensor may be faulty.
Possible
Solutions
Verify the physical offset.
Verify the pipe height.
Verify electronic offset and adjust if necessary.
Clean the sensor.
Replace the sensor.
Contact your regional ADS representative.