Maintenance and Troubleshooting 11-57
Range from the ultrasonic depth subsystem is too short
(but not zero).
Pipe height may be incorrect.
Electronic offset may be incorrect.
Verify electronic offset and adjust if necessary.
Contact your regional ADS representative.
Raw ultrasonic depth data shows depths greater than
the pipe diameter.
Physical offset may be incorrect.
Pipe height may be incorrect.
Electronic offset may be incorrect.
Verify the physical offset.
Verify electronic offset and adjust if necessary.
Contact your regional ADS representative.