Maintenance and Troubleshooting 11-59
Individual ultrasonic depth readings (SDEPTH_3) from
ParaDepth, ParaFlow,
Surface Combo
rasonic Depth Sensor are consistently
equivalent to the value obtained from the following
equation: Pipe Height
– Physical Offset + 9.99.
-volt (blue) fuse on regulator may have blown.
Replace the Surface Combo Sensor/Ultrasonic Depth
Sensor.
-volt (blue) fuse on the regulator.
:
Do not connect a potentially shorted sensor to
another monitor until confirming that the sensor is
functional (i.e., not shorted) using a voltmeter. A shorted
sensor will blow the 5
-
volt (blue) replacement fuse on
the regulator of the monitor.
Sensor reports abnormal temperatures.
Sensor connection to monitor may be loose.
Temperature sensor may be faulty.
Secure sensor connection to monitor if necessary.