11-64 ADS TRITON+ Manual
Problem
Velocity data does not fluctuate much (but is not 0).
Possible
Causes
Sensor may be dirty.
Minimum flows may be occurring within the sensor
deadband (extending 1.0 inches (25 mm) up from the
bottom of the Peak Combo Sensor).
Velocity function may be broken.
Possible
Solutions
Clean the sensor’s transmit/receive surfaces.
Verify that minimum flows are occurring outside the
sensor
deadband.
Check the sensor connections.
Check the velocity parameters.
Contact your regional ADS representative.
Replace the velocity sensor.