A-4 ADS TRITON+ Manual
Certified under the ATEX European Intrinsic
Safety standards for Zone 0 rated hazardous
areas
. Sira 09ATEX2027X
Certified under IECEx (International
Electrotechnical Commission Explosion Proof)
Intrinsic Safety Standards for use
in Zone 0
(equivalent to Class I, Division 1, Groups C & D)
rated hazardous areas
. IECEx SIR 09.0020X
CSA Certified to Class 225803 – Process Control
Equipment, Intrinsically Safe and Non
-Incendive
Systems
– For Zone 0 Hazardous Locations, Ex
ia IIB T3
Ga (152°C) in Canada. CSA
13CA2671180X
Certified to Class 225883 – Process Control
Equipment, Intrinsically Safe and Non-Incendive
Systems
– For Class I Zone 0 Hazardous
Locations, AEx ia IIB T3 Ga (152°C) in the USA
(equivalent to Class I, Division 1, Groups C & D).
CSA
13CA2671180X
Certified under UKCA (United Kingdom
Conformity Assessment) Intrinsic Safety
standards for Zone 0 Hazardous areas. CSAE
UKCA 22UKEX1159X
Calculating the Total Days of Storage Available
This section includes a special equation to use in determining the
number of days of data that can be stored in the monitor memory
based on a specific application. Consider the following when
calculating the total days:
x The monitor memory holds 832 kilobytes (or 851,968 bytes) of
data.
x The log rate determines the number of readings taken per day.
At a 15-minute sample rate, the monitor will record 96 readings
a day for a single entity. At a 5-minute sample rate, the
monitor will record 288 readings a day for a single entity.
x One entity reading consumes 8 bytes of monitor memory.
x The following table lists the standard number of entities
recorded for each sensor device. The date/time stamp for a