First Time Operation
Chapter
7
Operating (top) electrode:
Connect the 100Ω resistor between the Operating (top) electrode and the Common
(lower) electrode.
1. Press .
2.
H###
is displayed (
###
represents the calibration value).
3. Press a second time to calibrate the electrode.
AdJt
appears.
4. The electrodes have been calibrated.
5. Press to exit editing mode.
Reviewing Calibrated Values of Electrodes
If milk meter fault leads to suspecting calibration problem, the technician may check
calibration value at the display:
1. Ensure that Afimilk MPC is in wash mode
(
wash LED illuminated ).
2. Type
2580
and press to access programming mode.
3. Press Group access number -
4
to enter Calibration Group
4. Scroll down to
oA
(Operating Electrode).
6.
oA##
appears on the display panel. (
##
represents the calibration value.) If the
resulting value is less than 17 or more than 34, electrode calibration is required.
7. Press to scroll to calibration value
oH
(Holding Electrode).
oH##
appears on the display panel. (
##
represents the calibration value.) If the
resulting value is less than 17 or more than 33, electrode calibration is required.
5. Press + (double press) to return to wash mode.